Carnitine / Acylcarnitines LC-MS/MS Analysis Kit (Dried Blood Spot)
|
|
- Oz Aşık
- 6 yıl önce
- İzleme sayısı:
Transkript
1 Carnitine / Acylcarnitines LC-MS/MS Analysis Kit (Dried Blood Spot) ZV C User Manual
2 Table Of Contents Page 1. INTENDED USE: SUMMARY AND EXPLANATION: TEST PRINCIPLE: WARNING AND PRECAUTIONS: STORAGE AND STABILITY: MATERIALS SUPPLIED: MATERIALS REQUIRED BUT NOT SUPPLIED: PROCEDURE NOTES: LIMITATIONS OF THE PROCEDURE: PRE-TEST SET-UP INSTRUCTIONS: TEST PROCEDURE: QUALITY CONTROL: CALCULATION OF RESULTS: INTERPRETATION OF RESULTS: EXPECTED VALUES: LC-MS/MS PARAMETERS: MS SCAN PARAMETERS: ANALYTICAL PERFORMANCE: SAMPLE CHROMATOGRAM:
3 1. INTENDED USE: Carnitine / Acylcarnitines LC-MS/MS analysis kit for dried blood spot. 2. SUMMARY AND EXPLANATION: Carnitine / Acylcarnitines screening is the most important part of the clinical diagnosis of amino acid, fatty acid or organic acid metabolism. Carnitine / Acylcarnitines screening allows to the diagnosis of over 30 metabolic diseases from a drop of blood samples taken with a screening card. This screening test occurred by measurement of Carnitine / Acylcarnitins with isotope dilution tandem mass spectrometry. LC-MS/MS is a powerful tool for gives results Carnitine / Acylcarnitines in 1.5 minutes. Screening of following analytes can be performed by Zivak Carnitine / Acylcarnitine LC- MS/MS Analysis Kit and Zivak NeoZin Screening software. Main methods and procedures that have been selected are based on EN ISO By the quantitative analysis of carnitine esters including such as free carnitine, C0, C2, C4, C6, C8, C10, C12, C14, C16, C16 OH, C18, C18 OH carnitine disorders which are related with fatty acid oxidation. By the quantitative analysis of carnitine esters including such as C3, C3DC, C4OH, C5, C5OH,C5DC carnitine, disorders which are related with organic acid metabolism. 3. TEST PRINCIPLE: Carnitine / Acylcarnitines are extracted from dried blood spot with an organic solvent contains internal standards. Then extracted analytes derivatised via reagents. Derivatisated Carnitine / Acylcarnitines are analysed by Tandem Mass spectrometry. 3
4 4. WARNING AND PRECAUTIONS: For in-vitro diagnostic use only. For professional use only. Before you start, read the instructions carefully. In case of damage of the kit package, please contact Zivak or your supplier. Do not use expired kits and components. Please check batch no and expiry date before you start. Protective gloves and goggles should be worn. When working with biological fluids, please take any necessary precautions to prevent infection with blood borne pathogens. Appropriate bio-safety precautions and disposal of bio hazardous wastes should be followed Please check the labels on reagent bottles. Reagents of this kit contains hazardous material may cause eye and skin irritations. All calibrators and controls of this kit which contain human blood or dried blood was tested and found negative for HIV 1/2 and HCV antibodies, HbsAg, HIV 1/2 and HCV genome. Nevertheless, the blood controls should be considered as potentially infectious and treated with appropriate care. It s recommended to work in a fume hood for your safety as Reagent 2 is volatile and please avoid inhaling. 5. STORAGE AND STABILITY: This analysis kit can be shipped at room temperature. Controls, Reagent 2, 3 should be stored at 2-8 C. Reagent 1 should be stored at -20 C. All other components of the kit can be stored at room temperature. All components are guaranteed until expiry date when stored at recommended temperatures and used as described in these instructions. 6. MATERIALS SUPPLIED: Order No. Volume Symbol Component ZV R1-15 R1 1 x 40 ml 4 Reagent 1, Contains deuterated internal standards
5 ZV R x 12 ml R2 Reagent 2, Contains derivatisation reagent ZV R x 40 ml R3 Reagent 3, Contains acid ZV WB-15 1 x 0,5 L WB Washing solution, Contains organic solvent ZV MP-15 1 x 0,6 L MP Mobile Phase, Contains organic solvent KK ZV-3051-KK-15 1 x 1 pc User Manual 7. MATERIALS REQUIRED BUT NOT SUPPLIED: ZV K x 1 Spot Control Level 1 Dried Blood Spot Control Level 1, Contains Human Blood ZV K x 1 Spot Control Level 2 II Dried Blood Spot Control Level 2, Contains Human Blood ZV S x 1 Spot Calibrator Level 1 Dried Blood Spot Calibrator Level 1, Contains Human Blood ZV S x 1 Spot Calibrator Level 2 Dried Blood Spot Calibrator Level 2, Contains Human Blood Zivak Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Trap Column Housing Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Trap Column Cartridge µl pipette 5
6 Pipette tips 3.5 mm puncher Conical autosampler vial or vial insert User manual 8. PROCEDURE NOTES: Any improper handling of samples or modification of the test procedure may influence the results. The indicated pipetting volumes, incubation times, temperatures and pretreatment steps have to be performed strictly according to the instructions. Use calibrated pipettes and devices only. Once the test is started, all steps should be completed without interruption. Make sure that required reagents, materials and devices are prepared ready at the appropriate time. Allow all reagents and specimens to reach room temperature (18-25 C) and gently swirl each vial of liquid reagent and sample before use. Mix reagents without foaming. Avoid contamination of reagents, pipettes and wells/tubes. Use new disposable plastic pipette tips for each reagent, standard or specimen. Do not interchange caps. Always cap not used vials. Do not reuse wells/tubes or reagents. Incubation time affects results. All tubes or wells should be handled in the same order and time sequences. 6
7 9. LIMITATIONS OF THE PROCEDURE: Specimen collection has a significant effect on the test results. Please always follow the universal safety precautions for accurate and safe DBS sampling given below. Universal Safety Precautions for DBS sampling: Treat all blood samples as though they are infectious Wash hands, wear gloves and apron/lab coat Take precaution to avoid needle injury Dispose of contaminated sharps and waste appropriately Clearly label each card with appropriate identification number. It is unacceptable to submit a blood card for testing that has not been properly labelled. Apply gentle pressure to the heel and allow a large drop of free flowing blood to collect at the puncture site. Working quickly, hold the filter paper by the edges and touch the filter paper gently against the large drop of blood and in one step allow a sufficient quantity of blood to soak through and completely fill or saturate a circle. A completed saturated spot will contain approx.100 µl of blood. 7
8 Repeat, until you have collected enough blood to fill at least 3 circles on the blood collection card. It is critical that entire circle be uniformly saturated with blood. If collecting spots using a pipette, collect 100 µl of blood and gently apply to filter paper. Due to it is a screening method any result with an elevated concentration has to be indicated as presumptive positive and has to be confirmed with further sampling and confirmatory testing. A false negative result of this assay cannot be excluded with absolute certainty. 10. PRE-TEST SET-UP INSTRUCTIONS: Set-up the Instrument: Purge the HPLC pumps with a high flow rate of mobile phase(s). This should be done by pumping the mobile phase(s) through the system for 2 minutes at a flow rate of 4.0 ml/min. Switch off the pump and connect the column in flow direction. Activate the method and allow mobile phase(s) to flow through the pre-filter for 5 minutes. Make sure the bottle of mobile phase bottle is closed well, otherwise components of the mobile phase could evaporate; this alters the retention times. 8
9 11. TEST PROCEDURE: Sample Pre-treatment (Manual) Take dried blood spot sample in 3.5 mm diameter by a puncher on a 96-well microplate. Add 200 L of R1 to each well. Incubate for 30 minutes at room temperature. After incubation, take the liquid phase on another well. Then dry with warm air flow. Add 60 L of R2 on dried samples. Cover microplate properly with alumina foil or appropriate seal And incubate for 20 minutes at 65 C. Remove alumina foil and dry the liquid phase completely with warm air flow. Add 200 µl of R3 to dissolve the dried wells. Take the liquid into a vial or directly inject 20 µl from 96 well plate to LC-MS/MS system. Note: The prepared sample is stable at 2-8 C for 24 hours. 9
10 12. QUALITY CONTROL: The test results are only valid if the test is performed by the following the instructions. Moreover, the user must strictly adhere to the rules of GLP (Good Laboratory Practice) or other applicable standards/laws. Kit controls have to be found within the acceptable ranges as stated on the QC (Quality Control) Certificate. If the criteria are not met, the run is not valid and should be repeated. Each laboratory should use known samples as further controls. In case of any deviation, the following technical issues should be proven: Expiration dates of (prepared) reagents, storage conditions, pipettes, devices, incubation conditions and washing methods. It is recommended to participate at appropriate quality assessment trials. 13. CALCULATION OF RESULTS: Isotop dilution method is used for calculation of results. The analyte concentrations calculated against area of their internal standard with known concentrations. Using Zivak NeoZin Data Evaluation Software is very easy and effective software to analyse complex screening data. 10
11 14. INTERPRETATION OF RESULTS: Various societies for Carnitine / Acylcarnitines screening recommend different Cut-Off values for repetition of the measurement and the application of confirmatory assays. Depending on the application of samples of different populations of newborns it is highly recommended that each laboratory establishes its own range of normal values and that this distribution of values is co-ordinated with the recommendations of the responsible society of this geographic region. The results themselves should not be the only reason for any therapeutic consequences. They have to be correlated to other clinical observations and diagnostic tests. 15. EXPECTED VALUES: Expected values are set in the Zivak NeoZin Data Evaluation Software. It is recommended that each laboratory establishes its own range of normal values. 11
12 16. LC-MS/MS PARAMETERS: Device Column Injection Volume 20 µl Pump Program Flow Zivak Tandem Gold LC-MS/MS System Zivak Carnitine / Acylcarnitines Screening Dried Blood Spots LC- MS/MS Trap Column Housing - Cartridge 00:00 min 100% B 01:30 min 100% B 0.30 ml/min Ionization Mode: CID Gas API Nebulizing Gas Scan Time SIM Width Drying Gas ESI positive 2.0 mtorr 55 psi sec 1.5 amu 30 psi Drying Gas Temp. 350 C Needle Shield Capillary Detector 5000V 600V 30V 1500 V 12
13 17. MS SCAN PARAMETERS: No Analyte MH+ (m/z) MS/MS Capillary (ev) CE (ev) 1 C0 Carnitine C0 Carnitine-D C2 Carnitine C2 Carnitine-D C3 Carnitine C3DC Carnitine C3 Carnitine-D C4 Carnitine C4OH Carnitine C4 Carnitine-D C5 Carnitine C5DC Carnitine C5OH Carnitine C5 Carnitine-D C6 Carnitine C8 Carnitine C8 Carnitine-D C10 Carnitine C12 Carnitine C14 Carnitine C14 Carnitine-D C16 Carnitine C16OH Carnitine C16 Carnitine-D C18 Carnitine C18OH Carnitine Zivak Carnitine / Acylcarnitines Screening Dried Blood LC-MS/MS Analysis kit contains above analytes in controls. 13
14 INTERNAL STANDARD CONCENTRATIONS Analyte Final Concentration (µmol/l) C0 Carnitine-D C2 Carnitine-D C3 Carnitine-D C4 Carnitine-D C5 Carnitine-D C8 Carnitine-D C14 Carnitine-D C16 Carnitine-D
15 18. ANALYTICAL PERFORMANCE: No Analyte LOD (µmol/l) LOQ (µmol/l) Accuracy (%) Intra-Assay Precision (%CV) Inter-Assay Precision (%CV) Analytical Specificity (Cross Reactivity): No cross-reactivities were found with the typical substances tested. Linearity (R 2 ) 1 C0 Carnitine C2 Carnitine C3 Carnitine C4 Carnitine C5 Carnitine C4OH Carnitine C6 Carnitine C5OH Carnitine C8 Carnitine C3DC Carnitine C10 Carnitine C5DC Carnitine C12 Carnitine C16OH Carnitine C14 Carnitine C16 Carnitine C18OH Carnitine C18 Carnitine
16 19. SAMPLE CHROMATOGRAM: for Carnitine / Acylcarnitines 16
17 Sample chromatograms are taken from Carnitine / Acylcarnitines Screening Neozin Data Evaluation Software. 17
18 ACYL CARNITINES: C0-Carnitine: 18
19 Karnitin / Açil Karnitinler LC-MS/MS Analiz Kiti (Kuru Kanda) ZV C Kullanma Kılavuzu 19
20 İÇERİK Sayfa 1. KULLANIM AMACI: ÖZET VE AÇIKLAMA: TEST PRENSİBİ UYARILAR VE ÖNLEMLER SAKLAMA KOŞULLARI VE KARARLILIK KİT İÇERİĞİ KİT İÇERİĞİ DIŞINDA GEREKLİ OLAN MALZEMELER ÇALIŞMA PROSEDÜRÜ YÖNTEMDEKİ KISITLAMALAR ENSTRÜMAN ÖN AYARLARI NUMUNE HAZIRLIĞI KALİTE KONTROL SONUÇLARIN HESAPLANMASI SONUÇLARIN YORUMLANMASI BEKLENEN DEĞERLER MS/MS PARAMETRELERİ MS SCAN PARAMETRELERİ ÖRNEK KROMATOGRAM
21 1. KULLANIM AMACI: Kuru kan örneklerinde Karnitine / Açil Karnitinler tarama için LC-MS/MS analiz kiti. 2. ÖZET VE AÇIKLAMA: Karnitine / Açil Karnitinler tarama, aminoasit, yağ asitleri ve organik asit metabolizmalarındaki bozuklukların klinik tanısında en büyük rolü oynar. Karnitine / Açil Karnitinler tarama ile 30 dan fazla metabolik hastalığın teşhisini yapmak mümkündür. Bunun için bir damla kan örneği yeterlidir. Bu test bazı Karnitine / Açil Karnitinlerin izotop seyreltme yöntemine dayanarak tandem kütle spektrometrisi ile yapılır. LC-MS/MS sistemi Karnitine / Açil Karnitinler tarama için Zivak Karnitine / Açil Karnitinler Tarama Analiz kiti ile bütün bu analitlerin sonucunu 1.5 dakika içinde verebilen çok güçlü bir sistemdir. Aşağıdaki analitlerin tamamı Zivak Karnitine / Açil Karnitinler LC-MS/MS Analiz Kiti ve Zivak NeoZin software ile yapılabilir. Yağ asidi oksidasyonunda rol oynayan serbest karnitinler ve karnitin esterleri: C0, C2, C4, C6, C8, C10, C12, C14, C16, C16 OH, C18, C18 OH karnitin. Organik asit metabolizmasını rol oynayan karnitin esterler: C3, C3DC, C4OH, C5, C5OH,C5DC karnitin.. 21
22 3. TEST PRENSİBİ Karnitine / Açil Karnitinler organik çözücü içeren internal standart aracılığı ile kuru kandan ekstrakte edilir. Ekstrakte edilen analitler reaktifler ile türevlendirilir. Türevlendirilmiş Karnitine / Açil Karnitinler LC-MS/MS sisteminde analiz edilir. 4. UYARILAR VE ÖNLEMLER Sadece in-vitro kullanım amaçlıdır. Profesyonel kullanım içindir. Çalışmaya başlamadan önce, kullanım kılavuzunu dikkatlice okuyunuz. Kit ambalajının zarar görmüş olması durumunda, lütfen Zivak veya bayinizle irtibata geçiniz. Son kullanma tarihi geçmiş kitleri veya bileşenlerini kullanmayınız. Kullanmadan önce seri numarasını ve son kullanma tarihini kontrol ediniz. Toksik bileşen konsantrasyonları çok düşük olmakla birlikte güvenlik amacıyla çeker ocak altında çalışılması önerilir. Çalışırken koruyucu eldiven ve gözlük kullanılmalıdır. Çalışılan kan veya biyolojik örneklerden kaynaklanabilecek enfeksiyonlara karşı gerekli her türlü önlemi alınız. Biyo-zararlı atıklar için uygun biyo-güvenlik ve atık yönetimi koşullarını sağlayınız. Lütfen kullanmadan önce reaktif şişeleri üzerindeki etiketleri kontrol ediniz. Bu kitin reaktifleri göze ve cilde zarar verecek maddeler içerir. 5. SAKLAMA KOŞULLARI VE KARARLILIK R1 (-20) C de saklanmalıdır. R2, R3 ve kontroller 2-8 C de muhafaza edilebilir. Kitin diğer bileşenleri oda sıcaklığında saklanabilir. Bütün bileşenler önerilen koşullarda saklandığı takdirde son kullanma tarihine kadar garantilidir.. 22
23 6. KİT İÇERİĞİ Sipariş No Hacim Sembol İçerik ZV R x 40 ml R1 Reaktif 1, ZV R x 12 ml R2 Reaktif 2, ZV R x 40 ml R3 Reaktif 3, ZV WB-15 1 x 0,5 L WB Yıkama Solüsyonu ZV MP-15 1 x 0,6 L MP Mobil Faz KK ZV-3051-KK-15 1 x 1 ad Kullanma Klavuzu 23
24 7. KİT İÇERİĞİ DIŞINDA GEREKLİ OLAN MALZEMELER ZV K x 1 Spot Control Level 1 Kurukan Spotu Kontrol Level 1, ZV K x 1 Spot Control Level 2 II Kurukan Spotu Kontrol Level 2, ZV S x 1 Spot Calibrator Level 1 Kurukan Spotu Kalibrator Level 1, ZV S x 1 Spot Calibrator Level 2 Kurukan Spotu Kalibrator Level 2, KİT İÇERİĞİ DIŞINDA GEREKLİ OLAN MALZEMELER Zivak Karnitine / Açil Karnitinler Guard Kolon ve kartuş µl pipet Pipet uçları 3,5mm puncher Konik autosampler viali veya vial insert Kullanım Kılavuzu 8. ÇALIŞMA PROSEDÜRÜ Test prosedürünün yanlış ve değiştirilerek uygulanması test sonucunu etkileyebilir. Pipetleme hacimleri, inkübasyon zamanları ve hazırlık aşamaları tam olarak 24
25 belirtilen miktarlarda ve talimatlara uygun bir biçimde yapılmalıdır. Her zaman ve sadece kalibre edilmiş pipetler ve cihazlar kullanılmalıdır. Teste bir kere başlandığı zaman, herhangi bir kesintiye uğramadan tamamlanmalıdır. Bunu sağlamak için gerekli reaktiflerin, maddelerin ve cihazların zamanında hazır olduğundan emin olunuz. Çalışmaya başlamadan önce bütün reaktiflerin ve numunlerin oda sıcaklığına (18-25 C) gelmesini bekleyiniz, sıvı reaktifleri ve örneklerini kullanmadan önce hafifçe çalkalayarak karıştırınız. Reaktifler, pipetler ve numune hazırlık tüplerinde kirlilik ve bulaşma olmamasına dikkat ediniz. Her reaktif, standart ve numune için ayrı ve yeni pipet ucu kullanınız. Vial kapaklarını kendi aralarında değiştirmeyiniz. Kullanılmadığı zaman vial kapaklarını her zaman kapalı tutunuz. Vialler, vial kapakları, numune hazırlık tüpleri ve pipet uçları tek kullanımlıktır. İnkübasyon süreleri sonuçları etkilemektedir. Bütün numune hazırlık işlemleri belirtilen sırada ve sürede yapılmalıdır. 9. YÖNTEMDEKİ KISITLAMALAR Konsantrasyonalarda görülen herhangi bir artış tespit edildiğinde şüpheli pozitif olarak değerlendirilmeli ve tekrar numune alınıp test tekrar edilmelidir. Bu çalışmada herhangi yalancı negatif bir sonuç kesinlikle göz ardı edilemez. Numune toplama ve saklama koşulları test sonuçlarını belirgin bir şekilde etkilemektedir. Daha sonraki günlerde çalışılacak tam kan numuneleri analiz yapılana kadar dondurulmuş olarak saklanmalıdır. 10. ENSTRÜMAN ÖN AYARLARI. HPLC pompalarına mobil faz bağlandıktan sonra 4ml/dk akış ile 2 dakika boyunca purge yapılmalıdırlar. Metodu aktif ettikten sonar sistemin şartlanması için 15 dk beklenmelidir. 25
26 11. NUMUNE HAZIRLIĞI 1.Kurutulmuş kan örneğinden 3,5 mm çapında örnek alınarak 96-well microplate üzerinde bir kuyucuğa konulur. 3. İnkübasyondan sonra sıvı kısım başka bir kuyucuğa alınır. Sıcak hava akımı ile kurutulur. 4. f 2 kurutulmuş numunelerin üzerine eklenir. Mikroplate in üzeri alüminyum folyo ile iyice kapatılır. 20 dakika 65 C de bekletilir. 5. Alüminyum folyo çıkarılır. Kuyucuktaki sıvı sıcak hava akımı ile tamamen kurutulur. sistemine erek kurutulmuş kısım çözülür. 20 µl LC-MS/MS enjeksiyon yapılır. 12. KALİTE KONTROL Test sonuçları ancak verilen test talimatlarına uyulduğu takdirde geçerlidir. Kullanıcılar çalışırken GLP veya diğer kurumlar tarafından belirlenmiş olan kurallara ve standartlara bağlı kalmalıdır. 26
27 Bütün standart ve kontrol değerleri kalite kontrol sertifikasında belirtilen aralıklarda olmalıdır. Değerler bu aralıkların dışında kaldığı takdirde analiz geçerli değildir ve tekrarlanmalıdır. Her laboratuvar daha ileri aşama kontroller için değeri bilinen numuneleri kullanmalıdır. Bu değerlerden sapma olması halinde reaktiflerin son kullanma tarihi, saklama koşulları, kullanılan pipetlerin doğruluğu, cihazlar, inkübasyon koşulları ve yıkama metotları gibi teknik detayların doğruluğundan emin olunmalıdır. Ayrıca uygun dış kalite kontrol programına üye olunması önerilir. 13. SONUÇLARIN HESAPLANMASI Sonuçların hesaplanmasında izotop seyreltme metodu kullanılır. Analitlerin konsantrasyonları, konsantrasyonu bilinen internal standart alanlarına göre hesaplanır. Zivak NeoZin Newborn Screening Data Evaluation Software karmaşık tarama datalarını çok kolay ve etkili bir şekilde hesaplayan bir yazılımdır. 14. SONUÇLARIN YORUMLANMASI Farklı toplumlarda farklı Cut-Off değerleri belirlenmelidir. Farklı toplumlardaki hasta numuneleri için yapılan çalışmalara bağlı olarak, her laboratuvarın kendi normal değerlerini belirlemesi önemle tavsiye edilir ve bu değerlerin dağılımı o coğrafi bölgedeki sorumlu kurumun önerdiği değerler ile uyumlu olmalıdır. Tanı ve tedavi aşamalarında bu analizlerin sonuçları klinik gözlemler ve tanı testleri ile desteklenmelidir. 15. BEKLENEN DEĞERLER Her laboratuvarın kendi normal değerlerini belirlemesi önemle tavsiye edilir. 27
28 . 16. MS/MS PARAMETRELERİ Cihaz Zivak Tandem Gold LC-MS/MS Sistem Kolon Zivak- Karnitine / Açil Karnitinler Kolon Enjeksiyon Hacmi 20 µl Gradiyant dk 100% B 0,30 ml/dk 01:30 dk 100% B 0,30 ml/ dk Akış 0,30 ml/dk Iyonizasyon Modu ESI positive CID Gazı 2,0 mtorr API Nebulizing Gaz 55 psi Tarama Süresi 1,519 sn SIM Width 1,5 amu Drying Gaz 30 psi Drying Gaz.Sıc 350 C İğne Voltajı 5000V Shield Voltajı 600V Capillary 30V Dedektör
29 17. MS SCAN PARAMETRELERİ No Analit MH+ (m/z) MS/MS Capillary (ev) CE (ev) 1 C0 Carnitine C0 Carnitine-D C2 Carnitine C2 Carnitine-D C3 Carnitine C3DC Carnitine C3 Carnitine-D C4 Carnitine C4OH Carnitine C4 Carnitine-D C5 Carnitine C5DC Carnitine C5OH Carnitine C5 Carnitine-D C6 Carnitine C8 Carnitine C8 Carnitine-D C10 Carnitine C12 Carnitine C14 Carnitine C14 Carnitine-D C16 Carnitine C16OH Carnitine C16 Carnitine-D C18 Carnitine C18OH Carnitine Zivak Kurukanda Karnitine / Açil Karnitinler LC-MS/MS Analiz Kiti kontrolleri yukarıda listelenen analitleri içermektedir. 29
30 INTERNAL STANDART KONSANTRASYONLARI Final Konsantrasyonlar (µmol/l) C0 Carnitine-D9 1,52 C2 Carnitine-D3 0,38 C3 Carnitine-D3 0,076 C4 Carnitine-D3 0,076 C5 Carnitine-D9 0,076 C8 Carnitine-D3 0,076 C14 Carnitine-D9 0,076 C16 Carnitine-D3 0,152 30
31 18. ÖRNEK KROMATOGRAM Karnitinler için 31
32 32
33 33
34 C0-Carnitine: 34
Busulfan Serum & Plasma LC-MS/MS Analysis Kit
Busulfan Serum & Plasma LC-MS/MS Analysis Kit ZV-3011-0200-10 200 2-8 C User Manual Page 1. INTENDED USE:... 3 2. SUMMARY AND EXPLANATION:... 3 3. TEST PRINCIPLE... 3 4. WARNING AND PRECAUTIONS... 4 5.
DetaylıNeurotransmitters in Urine LC-MS/MS Analysis Kit
Neurotransmitters in Urine LC-MS/MS Analysis Kit ZV-3033-0200-20 200 2-8 C User Manual Table Of Contents Page 1. INTENDED USE:... 3 2. SUMMARY AND EXPLANATION:... 3 3. TEST PRINCIPLE:... 3 4. WARNING AND
Detaylı25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual
25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual ZV-4027-0500-10 500 2-8 C Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS... 4 4. TEST PRINCIPLE... 4 5.
Detaylı25-Hydroxyvitamin D2-D3 Serum LC-MS/MS APCI Analysis Kit User Manual ZV D. No: ZV-3046-KK-10.R15
25-Hydroxyvitamin D2-D3 Serum LC-MS/MS APCI Analysis Kit User Manual ZV-3046-0200-10 200 D. No: ZV-3046-KK-10.R15 Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS... 4 4. TEST
DetaylıELDAŞ Elektrik Elektronik Sanayi ve Tic.A.Ş.
Sayfa (Page): 2/9 LVD Deney Raporu LVD Testing Report İÇİNDEKİLER (Contents) 1 Dokümantasyon Sayfa (Documentation) 1.1 DGC, Çevre Koşulları ve Sembollerin Tanımları 3 (Conditions/Power Utilized,Description
DetaylıYüz Tanımaya Dayalı Uygulamalar. (Özet)
4 Yüz Tanımaya Dayalı Uygulamalar (Özet) Günümüzde, teknolojinin gelişmesi ile yüz tanımaya dayalı bir çok yöntem artık uygulama alanı bulabilmekte ve gittikçe de önem kazanmaktadır. Bir çok farklı uygulama
DetaylıGıda Analizlerinde LC-MS/MS Aplikasyonları
Gıda Analizlerinde LC-MS/MS Aplikasyonları Yem, Tahıl ve Yağlı tohum matrislerinde LC-MS/MS ile Mikotoksin ve Aflatoksin Kalıntısı miktarlandırma analizi, aplikasyon Çalışması. (Varian 320 MS Tandem Gold
DetaylıAKE Bulaşık Yıkama Makinası Kontrol Kartı Kullanım Kılavuzu Dishwasher Controller User Manual TR EN
Bulaşık Yıkama Makinası Kontrol Kartı Kullanım Kılavuzu Dishwasher Controller User Manual Bulaşık Yıkama Makinası Kontrol Kartı Kullanım Kılavuzu (7 SEG SIMPLE YATAY TİP) AKE-BYM-102 Lütfen bu kullanım
DetaylıH48.33b FILTER DRIER SHELLS KARTUŞ KOVANLARI. H48.33b.54.1
H48.33b H48.33b.54.1 Introduction Filter drier shells are important equipments for refrigeration and air conditioning systems. They are designed to be used in the suction line and liquid line. Filter drier
DetaylıNumune Kodu ve parti no
Numune Kodu ve parti no Numune tipi Kimin tarafından üretildiği Numune formu ve şekli Sertifikalandıran Kurum Konsantrasyonlar Elde edilen konsantrasyon değerleri Ortalama Standart Sapmalar % 95 Karbon
DetaylıÇEVRESEL TEST HİZMETLERİ 2.ENVIRONMENTAL TESTS
ÇEVRESEL TEST HİZMETLERİ 2.ENVIRONMENTAL TESTS Çevresel testler askeri ve sivil amaçlı kullanılan alt sistem ve sistemlerin ömür devirleri boyunca karşı karşıya kalabilecekleri doğal çevre şartlarına dirençlerini
DetaylıAKE.ZR Yıkama Durulama Zaman Rölesi Kullanım Kılavuzu Washing and Rinse Time Relay User Manual TR EN
Yıkama Durulama Zaman Rölesi Kullanım Kılavuzu Washing and Rinse Time Relay User Manual YIKAMA DURULAMA ZAMAN RÖLESİ Kullanım Kılavuzu Lütfen bu kullanım kılavuzunu cihazın montajını yapmadan ve cihazı
DetaylıUYGULAMA NOTU. LCMSMS ile Bebek Devam Formülleri ve Süt Tozunda Melamin Analizi. Sıvı Kromatografi Kütle Spektrometre HAZIRLAYAN
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M033 LCMSMS ile Bebek Devam Formülleri ve Süt Tozunda Melamin Analizi HAZIRLAYAN Dr. Engin Bayram Ant Teknik Cihazlar Ltd. Şti. KONU: Bebek devam formülü
DetaylıOrganic Acid Urine LC-MS/MS Analysis Kit
Organic Acid Urine LC-MS/MS Analysis Kit ZV-3004-0200-20 200 2-8 C User Manual D. No: ZV-3004-KK-20_Rev13 Table Of Contents Page 1. INTENDED USE:... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE:...
DetaylıKULLANICI TARAFINDAN TESTİN DOĞRULANMASI (VERİFİKASYON) Dr. Murat Öktem Düzen Laboratuvarlar Grubu
KULLANICI TARAFINDAN TESTİN DOĞRULANMASI (VERİFİKASYON) Dr. Murat Öktem Düzen Laboratuvarlar Grubu Kaynaklar CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods (2004)
DetaylıMETOT VALİDASYONU VE VERİFİKASYONU. Sedat Abuşoğlu Konya
METOT VALİDASYONU VE VERİFİKASYONU Sedat Abuşoğlu Konya İki Tanım Validasyon (CLSI): Bir sistem veya yöntemin beklendiği şekilde çalıştığını kanıtlama eylemi veya sürecidir. Validasyon (WHO-BS/95.1793):
DetaylıDelta Pulse 3 Montaj ve Çalıstırma Kılavuzu. www.teknolojiekibi.com
Delta Pulse 3 Montaj ve Çalıstırma Kılavuzu http:/// Bu kılavuz, montajı eksiksiz olarak yapılmış devrenin kontrolü ve çalıştırılması içindir. İçeriğinde montajı tamamlanmış devrede çalıştırma öncesinde
DetaylıWaspo ya Güvenebilirsiniz
Waspo, 30 yılı aşkın süredir otomotiv sektöründe faaliyet göstermekte olan Yeni Arçek İç ve Dış Ticaret LTD. ŞTİ. nin yeni markası olarak ticari hayatına başlamıştır. Kaliteli ürün, hızlı ve zamanında
DetaylıInspection Report Muayene Raporu KARE MÜHENDİSLİK ÇEVRE TEKNOLOJİLERİ SAN. TİC. A.Ş. SOLAR WATER HUMANITARIAN PROJECT
Offer Nr / Teklif No TGK Order Nr / Sipariş No Register Nr / Kayıt No 17-IS-0537-01-G 712917258 17 - IR - 0248.R1 Inspection Report Muayene Raporu No. KARE MÜHENDİSLİK ÇEVRE TEKNOLOJİLERİ SAN. TİC. A.Ş.
DetaylıQMS TOPIRAMAT. İmmün testi. Simgelerin Açıklamaları TOPIRAMATE KONTROLLERİ 0374181 1-3
0155241-C October, 2014 QMS TOPIRAMAT İmmün testi Bu Quantitative Microsphere System (QMS) prospektüsü kullanım öncesinde dikkatle okunmalıdır. Prospektüste verilen talimatlar uygulanmalıdır. Bu prospektüste
DetaylıMedical Waste. Kaya Süer MD. NEU Faculty of Medicine. Infectious Diseases and Clinical Microbiology
Medical Waste Kaya Süer MD. NEU Faculty of Medicine Infectious Diseases and Clinical Microbiology Household waste: Separately stored in black plastic bags. If mixing of medical waste shall be considered
DetaylıBVCD DÝKDÖRTGEN HAVA DAMPERÝ ( BVCD) RECTANGULAR VOLUME CONTROL DAMPER
DÝKDÖRTGEN HAVA DAMPERÝ ( ) RECTANGULAR VOLUME CONTROL DAMPER Dikdörtgen Hava Damperi Teknik Detaylarý Rectangular Volume Control Damper Technical Details : Malzeme: Ürün kasa ve kanatlarýnýn tamamý ekstrüzyon
DetaylıHIGH SPEED PVC DOOR INSTALLATION BOOK
HIGH SPEED PVC DOOR INSTALLATION BOOK HIZLI PVC KAPI MONTAJ KLAVUZU MODEL FUD 2015.01 MONTAJ KLAVUZU/INSTALLATION BOOK INTRODUCTION The information contained in this manual will allow you to install your
DetaylıCmpE 320 Spring 2008 Project #2 Evaluation Criteria
CmpE 320 Spring 2008 Project #2 Evaluation Criteria General The project was evaluated in terms of the following criteria: Correctness (55 points) See Correctness Evaluation below. Document (15 points)
DetaylıÖRNEKTİR - SAMPLE. RCSummer Ön Kayıt Formu Örneği - Sample Pre-Registration Form
RCSummer 2019 - Ön Kayıt Formu Örneği - Sample Pre-Registration Form BU FORM SADECE ÖN KAYIT FORMUDUR. Ön kaydınızın geçerli olması için formda verilen bilgilerin doğru olması gerekmektedir. Kontenjanımız
DetaylıAtıksu Arıtma Tesislerinde Hava Dağıtımının Optimize Edilmesi ve Enerji Tasarrufu
Optimization of Air Distribution in Waste Water Treatment Plants to Save Energy Atıksu Arıtma Tesislerinde Hava Dağıtımının Optimize Edilmesi ve Enerji Tasarrufu Jan Talkenberger, Binder Group, Ulm, Germany
Detaylı10.7442 g Na2HPO4.12H2O alınır, 500mL lik balonjojede hacim tamamlanır.
1-0,12 N 500 ml Na2HPO4 çözeltisi, Na2HPO4.12H2O kullanılarak nasıl hazırlanır? Bu çözeltiden alınan 1 ml lik bir kısım saf su ile 1000 ml ye seyreltiliyor. Son çözelti kaç Normaldir? Kaç ppm dir? % kaçlıktır?
DetaylıATILIM UNIVERSITY Department of Computer Engineering
ATILIM UNIVERSITY Department of Computer Engineering COMPE 350 Numerical Methods Fall, 2011 Instructor: Fügen Selbes Assistant: İsmail Onur Kaya Homework: 1 Due date: Nov 14, 2011 You are designing a spherical
DetaylıCalacatta Oro
Sayfa 1/8 Page 1/8 Müşterinin Adı/Adresi: Customer Name/Adress: Raport No: Report No: KOMMERSAN KOMBASSAN MERMER MADEN İŞLETMELERİ SAN VE TİC. A.Ş Muğla Aydın Karayolu 12. Km Salih Paşalar Mevkii Bayır
DetaylıPCC 6505 PROFILE CUTTING LINE
PCC 6505 PROFILE CUTTING LINE 1.DESCRIPTION PCC 6505 is a servo controlled machine which is specifically designed for the serial cutting of any kind of PVC and aluminum s in the market. The machine is
DetaylıOrganic Acid Urine LC-MS/MS Analysis Kit
Organic Acid Urine LC-MS/MS Analysis Kit ZV-3004-0200-20 200 2-8 C User Manual Table Of Contents Page 1. INTENDED USE:... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE:... 4 4. WARNING AND PRECAUTIONS:...
DetaylıQMS LAMOTRIGINE. İmmün testi. Simgelerin Açıklamaları LAMOTRİGİN KONTROLLERİ 0374090
0155206-D-TR October, 2014 QMS LAMOTRIGINE İmmün testi Bu Quantitative Microsphere System (QMS) prospektüsü kullanım öncesinde dikkatle okunmalıdır. Prospektüste verilen talimatlar uygulanmalıdır. Bu prospektüste
DetaylıCNC MACH breakout board user manual V8 type
CNC MACH breakout board user manual V8 type 1 Catalogue CNC Router breakout board V8 type user manual... Hata! Yer işareti tanımlanmamış. 1) Brief introduction:...3 2) Breakout board drawing:...4 3) Wiring:...5
DetaylıMezun ( ) Sınav Salon Numarası GENEL AÇIKLAMA (GENERAL INSTRUCTIONS) In the test,
Karadeniz Teknik Üniversitesi Yabancı Uyruklu Öğrenci Sınavı (KTÜYÖS) Karadeniz Technical University The Examination for Foreign Students A TEMEL ÖĞRENME BECERİLERİ TESTİ / THE BASIC LEARNING SKILLS TEST
DetaylıJeokimya Analizleri. Geochemical Analysis
Jeokimya Analizleri Geochemical Analysis Jeokimya; minerallerin oluşumu esnasında ve sonrasında çevre kayaçlar ile yüzeysel ortamlarda gerçekleşen kimyasal değişikliklerin belirlenmesi temeline dayanır.
DetaylıC6 Mugla White
Sayfa 1/8 Page 1/8 Müşterinin Adı/Adresi: Customer Name/Adress: Raport No: Report No: ÇOBANLAR MERMER MADENCİLİK TURZ. TİC. VE İNŞ. A.Ş. Salkım Mah. Çıngıl Sok. No:7 Kavaklıdere/MUĞLA 228 Numunenin Adı
DetaylıQMS LAMOTRIGINE. İmmün testi. Simgelerin Açıklamaları LAMOTRİGİN KALİBRATÖRLERİ 0373787
0155199-E-TR November, 2014 QMS LAMOTRIGINE İmmün testi Bu Quantitative Microsphere System (QMS) prospektüsü kullanım öncesinde dikkatle okunmalıdır. Prospektüste verilen talimatlar uygulanmalıdır. Bu
DetaylıC5 Mugla White
Sayfa 1/8 Page 1/8 Müşterinin Adı/Adresi: Customer Name/Adress: Raport No: Report No: ÇOBANLAR MERMER MADENCİLİK TURZ. TİC. VE İNŞ. A.Ş. Salkım Mah. Çıngıl Sok. No:7 Kavaklıdere/MUĞLA 227 Numunenin Adı
DetaylıVirtualmin'e Yeni Web Sitesi Host Etmek - Domain Eklemek
Yeni bir web sitesi tanımlamak, FTP ve Email ayarlarını ayarlamak için yapılması gerekenler Öncelikle Sol Menüden Create Virtual Server(Burdaki Virtual server ifadesi sizi yanıltmasın Reseller gibi düşünün
DetaylıUYGULAMA NOTU. LCMSMS Sistemi ile Tekstil, Deri ve Kağıt Materyallerinde Perflor Gruplu (PFOSs) Kimyasal Bileşiklerin Tayini
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M026 LCMSMS Sistemi ile Tekstil, Deri ve Kağıt Materyallerinde Perflor Gruplu (PFOSs) Kimyasal Bileşiklerin Tayini HAZIRLAYAN Yük. Kim. Murat ONUL Ant
DetaylıLIQUID RECEIVERS LİKİT TANKLARI VERTICAL LIQUID RECEIVERS DİKEY LİKİT TANKLARI HORIZONTAL & VERTICAL LIQUID RECEIVERS YATAY & DİKEY LİKİT TANKLARI
LIQUID RECEIVERS LİKİT TANKLARI VERTICAL LIQUID RECEIVERS HORIZONTAL & VERTICAL LIQUID RECEIVERS YATAY & DİKEY LİKİT TANKLARI LIQUID RECEIVERS LİKİT TANKLARI info@guvensogutma.com LIQUID RECEIVERS LİKİT
DetaylıFirst Stage of an Automated Content-Based Citation Analysis Study: Detection of Citation Sentences
First Stage of an Automated Content-Based Citation Analysis Study: Detection of Citation Sentences Zehra Taşkın, Umut Al & Umut Sezen {ztaskin, umutal, u.sezen}@hacettepe.edu.tr - 1 Plan Need for content-based
DetaylıMAIA Pesticide MultiTest
MAIA Pesticide MultiTest GIDALARDA PESTİSiT KALINTILARI İÇİN AB MAKSİMUM KALINTI LİMİTLERİ İLE UYUMLU ÇOKLU KALINTI TARAMA TESTİ Microplate Acetylcholinesterase Inhibition Assay (MAIA) katı veya sıvı gıda
DetaylıUYGULAMA NOTU. LCMSMS ile Bebek Mamasında Multitoksin Analizi - 1. Sıvı Kromatografi Kütle Spektrometre HAZIRLAYAN
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M030 LCMSMS ile Bebek Mamasında Multitoksin Analizi - 1 HAZIRLAYAN Dr. Engin Bayram Ant Teknik Cihazlar Ltd. Şti. KONU: Bebek mamasında multitoksinlerin
DetaylıTEST RESULTS UFED, XRY and SIMCON
TEST RESULTS UFED, XRY and SIMCON Test material : SIM card Tested software : UFED 3.6, XRY 6.5, SIMcon v1.2 Expected results : Proper extraction of SMS messages Date of the test : 02.04.2013 Note : The
DetaylıHAZIRLAYANLAR: K. ALBAYRAK, E. CİĞEROĞLU, M. İ. GÖKLER
HAZIRLAYANLAR: K. ALBAYRAK, E. CİĞEROĞLU, M. İ. GÖKLER PROGRAM OUTCOME 13 Ability to Take Societal, Environmental and Economical Considerations into Account in Professional Activities Program outcome 13
DetaylıBOXER. Contents TANITIM GÜVENLİK NAKİNEYİ KULLANMADAN ÖNCE, BU TALİMATLARI DİKKATLİCE OKUYUNUZ. UYARI
User manual Contents TANITIM Boks makinesi, bilgisayarla donatılmış, güç ölçmek için dizayn edilmiş bir makinedir. Boks makinesi, eğlence mekanlarında, oyun salonlarında ve eğlenceye yönelik mekanlarda
DetaylıUYGULAMA NOTU. LCMSMS ile Bebek Mamasında Multitoksin Analizi - 2. Sıvı Kromatografi Kütle Spektrometre HAZIRLAYAN
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M031 LCMSMS ile Bebek Mamasında Multitoksin Analizi - 2 HAZIRLAYAN Dr. Engin Bayram Ant Teknik Cihazlar Ltd. Şti. KONU: Bebek mamasında multitoksinlerin
DetaylıTeknoloji Servisleri; (Technology Services)
Antalya International University Teknoloji Servisleri; (Technology Services) Microsoft Ofis Yazılımları (Microsoft Office Software), How to Update Office 365 User Details How to forward email in Office
DetaylıRTA DNA qpcr Probe Master Mix
RTA DNA qpcr Probe Master Mix Kullanma Kılavuzu Yayın Tarihi -- 2012-01 DNA'nın gerçek zamanlı tayini ve miktar ölçümü için Yalnızca profesyonel kullanım için REF 09030025 25 test 09030100 100 test 09030500
DetaylıRTA JEL / PZR Saflaştırma Kiti
RTA JEL / PZR Saflaştırma Kiti Kullanma Kılavuzu Yayın Tarihi - 2011-12 DNA parçalarının agaroz jelden geri kazanımı ve PZR ürünlerinin saflaştırılması için Yalnızca profesyonel kullanım için REF 09009050
DetaylıUnlike analytical solutions, numerical methods have an error range. In addition to this
ERROR Unlike analytical solutions, numerical methods have an error range. In addition to this input data may have errors. There are 5 basis source of error: The Source of Error 1. Measuring Errors Data
Detaylızeytinist
1 T.C. BALIKESİR ÜNİVERSİTESİ EDREMİT MESLEK YÜKSEKOKULU Zeytincilik ve Zeytin İşleme Teknolojisi Programı Öğr. Gör. Mücahit KIVRAK 0 505 772 44 46 kivrak@gmail.com www.mucahitkivrak.com.tr 2 3 4 TEREYAĞI
DetaylıSirkülasyon Pompaları
Sirkülasyon Pompaları Circulation Pumps Dünyayı koruyan teknolojilerle daha konforlu bir hayat için çalışıyoruz.. Dünyanın önde gelen markaları, uzmanları, bilim insanları ve profesyonel kadroları, gelişen
DetaylıDersin Kodu Dersin Adı Dersin Türü Yıl Yarıyıl AKTS 507004092007 MAKİNA PROJESİ II Zorunlu 4 7 4
Ders Öğretim Planı Dersin Kodu Dersin Adı Dersin Türü Yıl Yarıyıl AKTS 507004092007 MAKİNA PROJESİ II Zorunlu 4 7 4 Dersin Seviyesi Lisans Dersin Amacı Dersin amacı Makina Mühendisliği bölümü Lisans öğrencilerine
DetaylıEge Üniversitesi Elektrik Elektronik Mühendisliği Bölümü Kontrol Sistemleri II Dersi Grup Adı: Sıvı Seviye Kontrol Deneyi.../..
Ege Üniversitesi Elektrik Elektronik Mühendisliği Bölümü Kontrol Sistemleri II Dersi Grup Adı: Sıvı Seviye Kontrol Deneyi.../../2015 KP Pompa akış sabiti 3.3 cm3/s/v DO1 Çıkış-1 in ağız çapı 0.635 cm DO2
DetaylıYeni Itru Lif Test Gereci UAK-1A
Yeni Itru Lif Test Gereci UAK-1A Teknik Bilgi http://www.itru.net E-mail :info@itru.net 1992-2004 by Itru Electronics Her hakkı saklıdır. TRU FIBRE TESTER UAK-1+ IX.yy da yaşamış TÜRK Matematikçisi (Musa
DetaylıCorporate Design Manual bio.inspecta Ltd. bio.inspecta Ltd. Kurumsal Tasarım El Kitabı
Corporate Design Manual bio.inspecta Ltd. bio.inspecta Ltd. Kurumsal Tasarım El Kitabı 13_121 Version 1.0 06.06.2016 1 Content/İçerik 1. Reference to certification body and logo on products/ Üründe sertifikasyon
DetaylıÖğrencilere, endüstriyel fanları ve kullanım alanlarını tanıtmak, endüstriyel fan teknolojisini öğretmektir.
Ders Öğretim Planı Dersin Kodu Dersin Adı Dersin Türü Yıl Yarıyıl AKTS 507004962014 ENDÜSTRİYEL FANLAR Seçmeli 4 8 3 Dersin Amacı Öğrencilere, endüstriyel fanları ve kullanım alanlarını tanıtmak, endüstriyel
DetaylıUYGULAMA NOTU. HPLC ile Gıda Ürünlerinde Fenolik Bileşen Analizi. Yüksek Performanslı Sıvı Kromatografi HAZIRLAYAN
UYGULAMA NOTU Yüksek Performanslı Sıvı Kromatografi L018 HPLC ile Gıda Ürünlerinde Fenolik Bileşen Analizi HAZIRLAYAN Uzm. Kim. Ozan Halisçelik ve Kim. Ömer H. Turmuş Ant Teknik Cihazlar Ltd. Şti. KONU:
DetaylıPOMPA VE MOTOR PUMPS AND MOTORS
8 tarım agriculture sanayi industry su idareleri water authority POMPA VE S AND S " 8" PASLANMAZ POMPA 8" SS 6 8 8 10 10 SS 877 KOMPLE PASLANMAZ POMPA FABRICATED STAINLESS STEEL 60 Hz 60 Hz seçim aralığı:
DetaylıDersin Kodu Dersin Adı Dersin Türü Yıl Yarıyıl AKTS
Dersin Kodu Dersin Adı Dersin Türü Yıl Yarıyıl AKTS 507004832007 KALİTE KONTROLÜ Seçmeli 4 7 3 Dersin Amacı Günümüz sanayisinin rekabet ortamında kalite kontrol gittikçe önem kazanan alanlardan birisi
DetaylıLIQUID RECEIVERS LİKİT TANKLARI VERTICAL LIQUID RECEIVERS DİKEY LİKİT TANKLARI HORIZONTAL & VERTICAL LIQUID RECEIVERS YATAY & DİKEY LİKİT TANKLARI
LIQUID RECEIVERS LİKİT TANKLARI VERTICAL LIQUID RECEIVERS DİKEY LİKİT TANKLARI HORIZONTAL & VERTICAL LIQUID RECEIVERS YATAY & DİKEY LİKİT TANKLARI LIQUID RECEIVERS LİKİT TANKLARI info@guvensogutma.com
DetaylıD-Link DSL 500G için ayarları
Celotex 4016 YAZILIM 80-8080-8081 İLDVR HARDWARE YAZILIM 80-4500-4600 DVR2000 25 FPS YAZILIM 5050-5555-1999-80 EX-3004 YAZILIM 5555 DVR 8008--9808 YAZILIM 80-9000-9001-9002 TE-203 VE TE-20316 SVDVR YAZILIM
DetaylıLCMSMS ile Gıdalarda Multitoksin Analizi
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M016 LCMSMS ile Gıdalarda Multitoksin Analizi HAZIRLAYAN Dr. Engin BAYRAM Ant Teknik Cihazlar Ltd. Şti. KONU: İnsanlar ve hayvanlar için yüksek toksisiteye
DetaylıSelçuk Üniversitesi Ziraat Fakultesi Bahçe Bitkileri Bolumu Selçuklu/KONYA (Sorumlu Yazar)
VII. Bahçe Ürünlerinde Muhafaza ve Pazarlama Sempozyumu, 04-07 Ekim 2016 ISSN: 2148-0036 Yıl /Year: 2017 Cilt(Sayı)/Vol.(Issue): 1(Özel) Sayfa/Page: 40-45 Araştırma Makalesi Research Article Selçuk Üniversitesi
DetaylıGENİŞLEYEN GÜVENLİK KAPISI EXPANDING SAFETY GATE
GENİŞLEYEN GÜVENLİK KAPISI EXPANDING SAFETY GATE www.odabu.com.tr UYARI MONTAJ VE KULLANIM ÖNCESİ KILAVUZ İÇİNDEKİ TÜM TALİMATLARI OKUYUNUZ KILAVUZU İHTİYAÇ DURUMUNDA KULLANMAK ÜZERE SAKLAYINIZ Güvenlik
Detaylı: Shower Unit (Flat) : Kompakt Duș Ünitesi (Flat)
Veo Description Tanım : Shower Unit (Flat) : Kompakt Duș Ünitesi (Flat) Left/Sol Right/Sağ Size / Ebat (cm) : 190x90 Depth / Derinlik (cm) : 3,5 Height / Yükseklik (cm) : 215 Weight / Ağırlık (kg) : min.
Detaylıeconn (Supplier Portal) of the MANN+HUMMEL Group
econn (Supplier Portal) of the MANN+HUMMEL Group October 2016 1 econn CONNECTED FOR SUCCESS Support If you have any questions or technical issues during the registration please contact: Teknik sorularınız
Detaylıverimli dayanıklı geniş ürün yelpazesi efficient durable wide range POMPA VE MOTOR PUMPS AND MOTORS
1 verimli dayanıklı geniş ürün yelpazesi efficient durable wide range POMPA VE S AND S 10" PASLANMAZ POMPA 10" SS 10" PASLANMAZ POMPA 10" SS 6 6 8 8 10 10 41 SS 10108 KOMPLE PASLANMAZ POMPA FABRICATED
DetaylıTÜM ÖĞRENCİ DEĞİŞİM PROGRAMLARI (ERASMUS-MEVLANA-FARABİ) BAŞVURU AŞAMALARI AYNI SÜRECİ TAKİP ETMEKTEDİR.
TÜM ÖĞRENCİ DEĞİŞİM PROGRAMLARI (ERASMUS-MEVLANA-FARABİ) BAŞVURU AŞAMALARI AYNI SÜRECİ TAKİP ETMEKTEDİR. ELİNİZDEKİ KLAVUZDA ÖRNEK OLARAK ERASMUS+ BAŞVURU SÜRECİ BELİRTİLMİŞTİR. ALL STUDENT WHO WILL APPLY
Detaylıverimli dayanıklı geniş ürün yelpazesi efficient durable wide range POMPA VE MOTOR PUMPS AND MOTORS
1 verimli dayanıklı geniş ürün yelpazesi efficient durable wide range POMPA VE S AND S 10" PASLANMAZ POMPA 10" SS 10" PASLANMAZ POMPA 10" SS 6 8 10 41 SS 10108 KOMPLE PASLANMAZ POMPA FABRICATED STAINLESS
DetaylıPRİZ KOLYE Clamp Saddles
PRİZ PR!Z KOLYE KOLYE C!VATALI!"#&'( (,' -%$.'/)!"#&'( (! *'01% "$'!.%3+1, /&%!.") #&4!.%)%,,((!"#&'( (,' -%$.'/) #&4!.%)%,,(( (!Ambalaj *'01% "$'!.%3+1, Adedi D x G /&%!.")!"#&'( 9111 x ½ 918 1 x ¾ 911
DetaylıYUVARLAK TAVAN ANEMOSTADI. Round Ceiling Diffuser
YUVARLAK TAVAN ANEMOSTADI Round Ceiling Diffuser YUVARLAK TAVAN ANEMOSTADI ( ) ROUND CEILING DIFFUSER Yuvarlak Tavan Anemostadý Teknik Detaylarý Round Ceiling Diffuser Technical Details : Malzeme: Kullaným:
DetaylıWEEK 11 CME323 NUMERIC ANALYSIS. Lect. Yasin ORTAKCI.
WEEK 11 CME323 NUMERIC ANALYSIS Lect. Yasin ORTAKCI yasinortakci@karabuk.edu.tr 2 INTERPOLATION Introduction A census of the population of the United States is taken every 10 years. The following table
DetaylıDo not open the exam until you are told that you may begin.
ÖRNEKTİR ÖRNEKTİR ÖRNEKTİR ÖRNEKTİR ÖRNEKTİR OKAN ÜNİVERSİTESİ FEN EDEBİYAT FAKÜLTESİ MATEMATİK BÖLÜMÜ 03.11.2011 MAT 461 Fonksiyonel Analiz I Ara Sınav N. Course ADI SOYADI ÖĞRENCİ NO İMZA Do not open
DetaylıUYGULAMA NOTU. LCMSMS ile Gıdalarda Sentetik Boyaların Analizi (Sudan Boyaları) Sıvı Kromatografi Kütle Spektrometre HAZIRLAYAN
UYGULAMA NOTU Sıvı Kromatografi Kütle Spektrometre M027 LCMSMS ile Gıdalarda Sentetik Boyaların Analizi (Sudan Boyaları) HAZIRLAYAN Dr. Engin Bayram Ant Teknik Cihazlar Ltd. Şti. KONU: Sudan Boyaları (SudanI,
DetaylıKULLANMA KLAVUZU / USER'S GUIDE GTİP/HS Code:
MOTORLU ÇOKLU ASTAR KESME MAKİNESİ MOTOR DRIVEN LINING CUTTING MACHINE KULLANMA KLAVUZU / USER'S GUIDE GTİP/HS Code: 8461.90.00.00.00 Çelik Kuyumculuk Kalıp Makine San. Ve Tic. Ltd. Şti. Adres: Atatürk
DetaylıDOKUZ EYLUL UNIVERSITY FACULTY OF ENGINEERING OFFICE OF THE DEAN COURSE / MODULE / BLOCK DETAILS ACADEMIC YEAR / SEMESTER
Offered by: Bilgisayar Mühendisliği Course Title: COMPUTER PROGRAMMING Course Org. Title: COMPUTER PROGRAMMING Course Level: Course Code: CME 0 Language of Instruction: İngilizce Form Submitting/Renewal
DetaylıMitsubishi Electric Corporation. Number of Items
Mitsubishi K65HM-CE / KP65HM-CE High Density Thermal Paper Monochrome High Density Thermal Sensitive Paper A6 type media 110mm x 21m Bluish tone 4 rolls per box For use in CP-700DSU, CP-700U, CP-700UM,
DetaylıGEO-Treat 2000 Series / Serisi. GEO-Treat 1000 Series / Serisi RESERVOIR TEMPERATURE HIGHER THAN 180 C RESERVOIR TEMPERATURE BETWEEN 120 C C
1000 Series / Serisi RESERVOIR TEMPERATURE HIGHER THAN 180 C Scale inhibitors showing high performance even at low dosages which are designed for wells with reservoir temperatures higher than 180 C and
Detaylıattomol apo B-100 quicktype
attomol apo B-100 quicktype İnsan apolipoprotein B-100 (apo B-3500 mutasyonu) gen inde 10708G>A geçiş tespitine yönelik kit Sadece in vitro diagnostik kullanım içindir! 20 tespit sipariş numarası: 1015
DetaylıInterest on request, samples were sent to our laboratory for lighting fixtures, physical examination and tests.the following results were obtained.
21/04/2015 Sayı/Issue: 14 04 15 SÖ 065/00 İlgi/Subject: Interest on request, samples were sent to our laboratory for lighting fixtures, physical examination and tests.the following results were obtained.
DetaylıLif Tutucu Menfez. Fiber Filter Grille
Lif Tutucu Menfez Fiber Filter Grille LÝF TUTUCU MENFEZ (BFFG) FIBER FILTER GRILLE Model: * BFFG : Lif tutucu menfez. Malzeme: Kullaným: Montaj: Sipariþ Kodu: Ürün kasasý ve 0,5 x Ø 0,3 mm gözenekli filtre
DetaylıKALEIDOSCOPES N.1. Solo Piano. Mehmet Okonşar
KALEIDOSCOPES N.1 Solo Piano Mehmet Okonşar Kaleidoscopes, bir temel ses dizisi üzerine kurulmuş ve bunların dönüşümlerini işleyen bir dizi yapıttan oluşmaktadır. Kullanılan bu temel ses dizisi, Alban
DetaylıTÜBİTAK ULUSAL METROLOJİ ENSTİTÜSÜ
Referans Malzeme Sertifikası Certificate of the Reference Material Sayfa 1 / 5 Malzemenin Adı Name of the Material Malzemenin Kodu Material Code Onay Tarihi Issue Date Son Revizyon Tarihi Last Revision
DetaylıDairesel grafik (veya dilimli pie chart circle graph diyagram, sektor grafiği) (İngilizce:"pie chart"), istatistik
DAİRESEL GRAFİK Dairesel grafik (veya dilimli diyagram, sektor grafiği) (İngilizce:"pie chart"), istatistik biliminde betimsel istatistik alanında kategorik (ya sırasal ölçekli ya da isimsel ölçekli) verileri
DetaylıArýza Giderme. Troubleshooting
Arýza Giderme Sorun Olasý Nedenler Giriþ Gerilimi düþük hata mesajý Þebeke giriþ gerilimi alt seviyenin altýnda geliyor Þebeke giriþ gerilimi tehlikeli derecede Yüksek geliyor Regülatör kontrol kartý hatasý
Detaylı1 - CANTEK köşe parçasını şekilde gösterildiği gibi yerleştirin;
1 - CANTEK köşe parçasını şekilde gösterildiği gibi yerleştirin; Insert CANTEK Corner Piece as shown.. 2 - CANTEK Köşe parçalarını 4 köşeye yerleştirilerek şekildeki gibi çerçeve elde edin. Kanal kesitinin
DetaylıIDENTITY MANAGEMENT FOR EXTERNAL USERS
1/11 Sürüm Numarası Değişiklik Tarihi Değişikliği Yapan Erman Ulusoy Açıklama İlk Sürüm IDENTITY MANAGEMENT FOR EXTERNAL USERS You can connect EXTERNAL Identity Management System (IDM) with https://selfservice.tai.com.tr/
DetaylıİYC MADENCİLİK SAN. VE TİC. LTD. ŞTİ. NE AİT MUĞLA - FETHİYE YÖRESİ BEJ TÜRÜ KİREÇTAŞININ FİZİKO-MEKANİK ANALİZ RAPORU
T.C. PAMUKKALE ÜNİVERSİTESİ MÜHENDİSLİK FAKÜLTESİ PAMUKKALE UNIVERSITY FACULTY OF ENGINEERING İYC MADENCİLİK SAN. VE TİC. LTD. ŞTİ. NE AİT MUĞLA - FETHİYE YÖRESİ BEJ TÜRÜ KİREÇTAŞININ FİZİKO-MEKANİK ANALİZ
Detaylıaquanit.com.tr Uygulama - Montaj Kılavuzu Application - Installation Manual
aquanit.com.tr Model / Multi 90X90 / 90X15 aquanit.com.tr Uygulama Adımları / Application Steps Uygulama Adımları / Application Steps Sifon şablon ölçüsü alınır Measure the floor siphon template Montaj
DetaylıINSTALLATION INSTRUCTIONS. York Prelude Part #: Read carefully. Wear protective gloves
INSTALLATION INSTRUCTIONS York Prelude Part #: 1720 Confidence from start to finish. Read carefully Wear protective gloves JACLO.COM PG 1 PARTS 1. Wall support 2. Screws and expansion bolts 3. Cover 4.
DetaylıBebek Hamağı Kullanım Kılavuzu Baby Hammock User Guide
Türk Patent Enstitüsü Tasarım Tescilli 5 Dakikada Kurulum Alman TÜV Rheinland Tarafından Güvenlik Onaylı Bebek Hamağı Kullanım Kılavuzu Baby Hammock User Guide Ürünün kurulum ve kullanımından önce kılavuzu
DetaylıTANIMLAR. Dr. Neriman AYDIN. Adnan Menderes Üniversitesi Tıp Fakültesi Tıbbi Mikrobiyoloji Anabilim Dalı
TANIMLAR Dr. Neriman AYDIN Adnan Menderes Üniversitesi Tıp Fakültesi Tıbbi Mikrobiyoloji Anabilim Dalı In-vitro tanı Sürekli doğru sonuç Amaca uygun Zamanında Uygun maliyet In-vitro tanı elemanları Kitler
DetaylıVİRAL MARKER PROGRAMI EKİM-2013 DÖNEMİ DEĞERLENDİRMESİ
VİRAL MARKER PROGRAMI EKİM-2013 DÖNEMİ DEĞERLENDİRMESİ Viral marker programında katılımcı sayısı 180 e ulaşmıştır. Bu artışla birlikte tüm çalışma yöntemleri için de katılımcı sayıları artmaktadır. Çalışma
DetaylıSeri kablo bağlantısında Windows95/98/ME'ten Windows 2000'e bağlantı Windows95/98/ME - NT4 bağlantısına çok benzer.
Seri kablo bağlantısında Windows95/98/ME'ten Windows 2000'e bağlantı Windows95/98/ME NT4 bağlantısına çok benzer. Direkt Kablo desteğini Windows95/98'e yükledikten sonra, Windows95 for Direct Cable Client
DetaylıIslington da Pratisyen Hekimliğinizi ziyaret ettiğinizde bir tercüman istemek. Getting an interpreter when you visit your GP practice in Islington
Islington da Pratisyen Hekimliğinizi ziyaret ettiğinizde bir tercüman istemek Getting an interpreter when you visit your GP practice in Islington Islington daki tüm Pratisyen Hekimlikler (GP) tercümanlık
Detaylı: ALG Müh. ve Tic. Ahmet Lütfü Gönül ALG Su Laboratuvarı Sok. No: 90/205 El-Sel Sitesi Alsancak İzmir
0805010096 0805010096 Cihazın Sahibi / Adresi Customer Name / Address İstek Numarası Order Number : ALG Müh. ve Tic. Ahmet Lütfü Gönül ALG Su Laboratuvarı 1420 Sok. No: 90/205 El-Sel Sitesi Alsancak İzmir
DetaylıHPLC ile Elma Suyunda HMF Analizi
UYGULAMA NOTU Yüksek Performanslı Sıvı Kromatografi L019 HPLC ile Elma Suyunda HMF Analizi HAZIRLAYANLAR Kim. Akın Osanmaz ve Uzm. Kim. Ozan Halisçelik Ant Teknik Cihazlar Ltd. Şti. KONU: Elma suyu numunelerinde,
Detaylı